Gentle blood aspiration and tube cushioning reduce pneumatic tube system interference in lactate dehydrogenase assays.

ANNALS OF CLINICAL BIOCHEMISTRY(2015)

引用 10|浏览1
暂无评分
摘要
BACKGROUND:Use of a hospital pneumatic tube system may be associated with measurement errors. METHODS:A venous blood sample was collected from 79 patients into a pair of lithium heparin tubes; one tube was sent to the laboratory by porter and the other was sent via the pneumatic tube system. Plasma lactate dehydrogenase concentrations were then assayed. RESULTS:Lactate dehydrogenase concentrations were overestimated (median bias: 18.8%) when evacuated vacuum lithium heparin tubes were sent by pneumatic tube system. This bias was reduced by bubble-wrapping the standard lithium heparin tube or using Monovette lithium heparin tubes in aspiration mode (median bias: +8.7% and -0.3%, respectively). CONCLUSIONS:Cushioning and aspiration-mode sampling may limit pneumatic tube system-associated overestimation of lactate dehydrogenase concentrations.
更多
查看译文
关键词
Enzymes,analytes,interference,pneumatic tube system,lactate dehydrogenase