Comparative performances analysis of neonatal ventilators

Italian journal of pediatrics(2015)

引用 4|浏览5
暂无评分
摘要
Background Mechanical ventilation is a therapeutic action for newborns with respiratory diseases but may have side effects. Correct equipment knowledge and training may limit human errors. We aimed to test different neonatal mechanical ventilators’ performances by an acquisition module (a commercial pressure sensor plus an isolated chamber and a dedicated software). Methods The differences (ΔP) between peak pressure values and end-expiration pressure were investigated for each ventilator. We focused on discrepancies among measured and imposed pressure data. A statistical analysis was performed. Results We investigated the measured/imposed ΔP relation. The ΔP do not reveal univocal trends related to ventilation setting parameters and the data distributions were non-Gaussian. Conclusions Measured ΔP represent a significant parameter in newborns’ ventilation, due to the typical small volumes. The investigated ventilators showed different tendencies. Therefore, a deep specific knowledge of the intensive care devices is mandatory for caregivers to correctly exploit their operating principles.
更多
查看译文
关键词
Mechanical ventilation,Neonatal pulmonary medicine,Respiratory technology