Effects of electrode array length on frequency-place mismatch and speech perception with cochlear implants.

AUDIOLOGY AND NEURO-OTOLOGY(2015)

引用 28|浏览9
暂无评分
摘要
Frequency-place mismatch often occurs after cochlear implantation, yet its effect on speech perception outcome remains unclear. In this article, we propose a method, based on cochlea imaging, to determine the cochlear place-frequency map. We evaluated the effect of frequency-place mismatch on speech perception outcome in subjects implanted with 3 different lengths of electrode arrays. A deeper insertion was responsible for a larger frequency-place mismatch and a decreased and delayed speech perception improvement by comparison with a shallower insertion, for which a similar but slighter effect was noticed. Our results support the notion that selecting an electrode array length adapted to each individual's cochlear anatomy may reduce frequency-place mismatch and thus improve speech perception outcome. (C) 2015 S. Karger AG, Basel
更多
查看译文
关键词
Pitch,Cochlear implant,Speech perception,Cochlear tonotopy