Cost of a recall of a single-center experience managing the Riata defibrillator lead.

The American Journal of Cardiology(2015)

引用 2|浏览2
暂无评分
摘要
Riata and Riata ST defibrillator leads (St. Jude Medical, Sylmar, California) were recalled in 2011 due to increased risk of insulation failure leading to externalized cables. Fluoroscopic screening can identify insulation failure, although the relation between mechanical failure and electrical failure is unclear. At the time of the recall, the University of Virginia developed a screening program, including fluoroscopic evaluation, education sessions, device interrogation, and remote monitoring for patients with this defibrillator lead. The aim of this study was to review the outcomes of the screening program, including costs, which were absorbed by our institution. Costs were calculated using Medicare reimbursement estimates. Fortyeight patients participated in the screening program. At initial screening, 31% were found to have evidence of insulation failure but electrical function was normal in all leads. The cost of this program was $35,358.72. The cost per diagnosis of mechanical lead failure was $2,357.25. During 2 years of follow-up, 1 patient experienced Riata lead electrical failure without fluoroscopic evidence of insulation failure. Patients were more likely to have a lead revision if there was evidence of insulation failure. Lead revisions occurred at the time of generator change in 88% of patients with insulation failure but in only 14% of patients with a fluoroscopically normal lead (p = 0.04). The cost of recall-related defibrillator lead revisions was $81,704.55. In conclusion, our Riata screening program added expense without clear benefit to patients. In fact, patients may have been put at more risk by undergoing defibrillator lead revisions based solely on the results of the fluoroscopic screening. (C) 2015 Elsevier Inc. All rights reserved.
更多
查看译文
关键词
riata defibrillator lead,cost,recall,single-center
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要