Commissioning of a new SeHCAT detector and comparison with an uncollimated gamma camera.

NUCLEAR MEDICINE COMMUNICATIONS(2014)

引用 2|浏览1
暂无评分
摘要
Measurements of SeHCAT (tauroselcholic [75selenium] acid) retention have been used to diagnose bile acid malabsorption for a number of years. In current UK practice the vast majority of centres calculate uptake using an uncollimated gamma camera. Because of ever-increasing demands on gamma camera time, a new probe' detector was designed, assembled and commissioned. To validate the system, nine patients were scanned at day 0 and day 7 with both the new probe detector and an uncollimated gamma camera. Commissioning results were largely in line with expectations. Spatial resolution (full-width 95% of maximum) at 1m was 36.6cm, the background count rate was 24.7cps and sensitivity at 1m was 720.8cps/MBq. The patient comparison study showed a mean absolute difference in retention measurements of 0.8% between the probe and uncollimated gamma camera, and SD of +/- 1.8%. The study demonstrated that it is possible to create a simple, reproducible SeHCAT measurement system using a commercially available scintillation detector. Retention results from the probe closely agreed with those from the uncollimated gamma camera.
更多
查看译文
关键词
bile acid malabsorption,gamma probe,SeHCAT