The diagnosis of silicone breast-implant rupture: clinical findings compared with findings at magnetic resonance imaging.

Lisbet Rosenkrantz Hölmich,Jon P Fryzek, Kim Kjøller, Vibeke Bro Breiting, Anna Jørgensen,Christen Krag,Joseph K McLaughlin

ANNALS OF PLASTIC SURGERY(2007)

引用 92|浏览1
暂无评分
摘要
The objective was to evaluate the usefulness of clinical examination in the evaluation of breast-implant integrity, using the diagnosis at magnetic resonance imaging (MRI) as the "gold standard." Fifty-five women with 109 implants underwent a breast examination either just before or shortly after an M RI examination. Twenty-four of 109 implants were clinically diagnosed with possible rupture or rupture. Eighteen of the 24 implants were ruptured according to the MRI examination (75%). Eighty-five implants were clinically classified as intact, and 43 of these were actually ruptured at MRI (51%). The sensitivity of the clinical examination for diagnosing rupture was thus 30% and the specificity 88%. The positive predictive value of a clinical diagnosis of rupture was 75%, and the negative predictive value was 49%. In this study, we found that when a clinical examination is used as the sole diagnostic tool to identify implant rupture, neither the sensitivity nor the specificity is acceptable.
更多
查看译文
关键词
breast augmentation,breast implants,implant rupture,silicone gel,adverse effects