Spurious elevation of glucose concentration during administration of high dose of ascorbic acid in a patient with type 2 diabetes on hemodialysis.

YONSEI MEDICAL JOURNAL(2013)

引用 11|浏览17
暂无评分
摘要
We describe herein a case of life-threatening hypoglycemia due to spurious elevation of glucose concentration during the administration of ascorbic acid in a type 2 diabetic patient. A 31-year-old female was admitted for proliferative diabetic retinopathy treatment and prescribed high dose ascorbic acid. During hospitalization, she suddenly lost her consciousness and her glucose concentration was 291 mg/dL, measured using self-monitoring blood glucose (SMBG) device, while venous blood glucose concentration was 12 mg/dL. After intravenous injection of 50% glucose solution, the patient became alert. We reasoned that glucose measurement by SMBG device was interfered by ascorbic acid. Physicians should be aware of this interference; high dose ascorbic acid may cause spurious elevation of glucose concentration when measuring with SMBG devices.