A comparison of leak compensation during pediatric noninvasive ventilation: a lung model study.

RESPIRATORY CARE(2014)

引用 22|浏览14
暂无评分
摘要
BACKGROUND: Ventilators used for noninvasive ventilation (NIV) must be able to synchronize in the presence of system leaks. We compared the ability of 7 ICU ventilators and 3 dedicated NIV ventilators to compensate for leaks during pediatric NW. METHODS: Using a lung simulator, we compared the Magnet Servo-i, Drager V500, Drager Carina, Covidien PB840, Respironics V60, Respironics Vision, GE Healthcare/Engstrom Carestation, Care Fusion Avea, Hamilton C3, and Hamilton G5 during increasing (n = 6) and decreasing leaks (n = 6). With a lung simulator we tested 4 leak levels (baseline [BL] 2-3 L/min, L1 5-6 L/min, L2 9-10 L/min, and L3 19-20 L/min); 3 patient weights (10, 20, and 30 kg); and 3 lung mechanics scenarios (normal, obstructive, and restrictive). The ventilator settings were NIV mode, pressure support of 10 cm H2O, and PEEP of 5 cm H2O. The synchronization rate (synchronized cycles/total simulated respirations) was recorded for each ventilator and each leak scenario. Synchronization was defined as triggering without auto-triggering, miss-triggering, delayed cycling, or premature cycling. RESULTS: The mean synchronization rate across all ventilators was 68 +/- 27% (range 23-96%) and marked differences existed between the ventilators (P < .001). Significant differences in synchronization rate were observed between the 10-kg model (mean 57 +/- 30%, range 17-93%), the 20-kg model (69 +/- 30%, 25-98%), and the 30-kg models (77 +/- 22%, 28-97%) (P < .001). The synchronization rate for the obstructive model (60 +/- 30%, 9-94%) was significantly different from the normal model (71 +/- 29%, 18-98%) and the restrictive model (72 +/- 28%, 23-98%) (P < .001). The PB840 and the C3 had synchronization rates over 90% overall across all body weights, all lung mechanic profiles, and all leak levels. CONCLUSIONS: Leak compensation in NIV for pediatric use can partially compensate for leaks, but varies widely among ventilators, patient weights, and lung mechanics.
更多
查看译文
关键词
leak compensation,pediatric noninvasive ventilation,acute care ventilator
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要