Percutaneous Transvenous Mitral Annuloplasty (PTMA) with the Viking device reduces pacing-induced mitral regurgitation.

Per Ola Kimblad,Jan Harnek,Anders Roijer, Carl Meurling, Johan Brandt, Jan Otto Solem

EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology(2005)

引用 30|浏览2
暂无评分
摘要
The new percutaneous mitral annuloplasty Viking device was evaluated in surviving sheep with pacing-induced mitral regurgitation.Twenty sheep were subjected to rapid ventricular pacing for one to three months, leading to cardiomyopathy and mitral regurgitation. Device implantation could be successfully performed in 11 of these animals after pacemaker treatment for 64+/-7 days. The device-related procedure time was 12+/-2 min. The mean follow-up time was 58+/-8 days after implantation of the device. Mitral annulus septo-lateral diameter was significantly reduced after insertion of the device, from 35+/-1 mm before implantation to 30+/-1 mm at the final follow up intracardiac echocardiography (P = 0.0097). The degree of mitral regurgitation (on a scale from 0 to 4) was 2.6+/-0.2 before device implantation and decreased to 0.8+/-0.2 after treatment (P = 0.0039), and the vena contracta was reduced from 7+/-0.4 mm to 3+/-0.8 mm (P = 0.0019). Angiography showed no signs of impairment of the coronary arteries. No thrombosis was observed.These results indicate that the septo-lateral diameter of the mitral annulus, and the degree of experimentally induced mitral regurgitation, can be significantly reduced with a percutaneous catheter technique in surviving sheep.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要