This paper presents a 3D semi-automatic prostate segmentation method for B-mode trans-rectal ultrasound (TRUS) images. Segmentation is based on prior knowledge of the prostate transversal section shape which is assumed to be a tapered ellipse. The approach consists of an initial untapering and warping of the image to make the shape of the prostate approximately elliptical. The prostate contour is found in the untapered, warped images by edge detection and 2D/3D ellipsoidal curve fitting, performed by solving convex optimization problem. This leads to an average total segmentation duration of 7.42s which is more than 40 times faster than manual segmentation. Clinical studies carried out on 58 prostate volume studies show a volume sensitivity and accuracy of approximately 94% and 73% respectively.