Assessment of patient dose and image quality for cardiac CT with breast shields.

RADIATION PROTECTION DOSIMETRY(2012)

引用 7|浏览6
暂无评分
摘要
Breast shielding can reduce dose to the female breast, a radiosensitive organ receiving significant radiation during computed tomography (CT) chest examinations, particularly in cardiac CT, where Electrocardiogram dose modulation currently precludes the use of radial dose modulation to reduce breast dose. However, breast shields may produce artefacts affecting interpretation of coronary arteries. This study explores the dose savings and the effect of breast shields on image quality with torso and CT dose index body phantoms and an organ dose calculator. Change in dose calculated: 5363 (female breast), 8285 (lung), 7984 (oesophagus) and 7680 (effective dose) with larger dose reductions at lower kVp. Image quality is preserved when breast shields are placed after the scout no closer than 10 mm from the skin. Therefore, breast shields can be used in cardiac CT to reduce breast dose without compromising image quality. Revised conversion factors for dose length product to effective dose are suggested for cardiac CT without and with breast shields.