Dead-end ultrafiltration concentration and IMS/ATP-bioluminescence detection of Escherichia coli O157:H7 in recreational water and produce wash.

Journal of Microbiological Methods(2011)

引用 18|浏览1
暂无评分
摘要
The purpose of this study was to develop a detection method for viable E. coli O157:H7 in fresh produce and recreational water. The method was evaluated using eight samples of produce wash and recreational water with or without spiked E. coli O157:H7 at ≤102CFU·ml−1 and concentrated using dead-end ultrafiltration (DEUF) to produce primary and secondary retentates. Fifty-four matrix replicates of undiluted secondary retentates or dilutions (1:2 or 1:10 in buffer) were evaluated using an IMS/ATP bioluminescence assay (IMS/ATP). Combining primary and secondary DEUF yielded a 2–4 log10 increase in E. coli O157:H7 concentrations in spiked samples and resulted in signal-to-noise ratios 2–219 fold higher than controls, depending on the sample type. DEUF increased the concentration of E. coli O157:H7 to within the detectable limits of IMS/ATP. The combined assay provided detection of viable E. coli O157:H7 in produce and recreational water. Accurate detection of microbial pathogens using DEUF and IMS/ATP could reduce disease outbreaks from contaminated water sources and food products.
更多
查看译文
关键词
ATP bioluminescence,IMS,E. coli O157:H7,Produce,Recreational water,Ultrafiltration