Safe neuromuscular electrical stimulator designed for the elderly.

ARTIFICIAL ORGANS(2011)

引用 25|浏览3
暂无评分
摘要
A stimulator for neuromuscular electrical stimulation (NMES) was designed, especially suiting the requirements of elderly people with reduced cognitive abilities and diminished fine motor skills. The aging of skeletal muscle is characterized by a progressive decline in muscle mass, force, and condition. Muscle training with NMES reduces the degradation process. The discussed system is intended for evoked muscle training of the anterior and posterior thigh. The core of the stimulator is based on a microcontroller with two modular output stages. The system has two charge-balanced biphasic voltage-controlled stimulation channels. Additionally, the evoked myoelectric signal (M-wave) and the myokinematic signal (surface acceleration) are measured. A central controller unit allows using the stimulator as a stand-alone device. To set up the training sequences and to evaluate the compliance data, a personal computer is connected to the stimulator via a universal serial bus. To help elderly people handle the stimulator by themselves, the user interface is kept very simple. For safety reasons, the electrode impedance is monitored during stimulation. A comprehensive compliance management with included measurements of muscle activity and stimulation intensity enables a scientific use of the stimulator in clinical trials.
更多
查看译文
关键词
Neuromuscular electrical stimulation,Aging muscle,Electrical stimulation,Compliance management,M-wave,Myokinematic signal