Acute Complication Due To Impella 2.5 Device (Superficial Femoral Artery Thrombosis): Managed Successfully With Novel Aspiration Thrombectomy Catheter (Pronto V3)

CLINICAL MEDICINE INSIGHTS-CARDIOLOGY(2011)

引用 9|浏览1
暂无评分
摘要
The Impella recover LP 2.5 is a percutaneous left ventricular assist device (LVAD) recently approved for use in patients undergoing high risk percutaneous coronary intervention (PCI) and also in cases of cardiogenic shock. There is limited evidence available in literature about its safety, especially with regards to the incidence of local vascular complications, their management and long-term implications. We report here the first case of a serious local vascular complication-superficial femoral artery thrombus formation during Impella recover LP 2.5 use in a high risk PCI which was managed successfully with novel aspiration thrombectomy catheter (Pronto V3), which in itself is the first reported use of Pronto V3 in such a vascular complication.
更多
查看译文
关键词
impella LP 2.5, complication of impella, pronto V3, high risk PCI