Accuracy Of Diagnoses Delivered By An Automated Hand-Held Nerve Conduction Device In Comparison To Standard Electrophysiological Testing In Patients With Unilateral Leg Symptoms

MUSCLE & NERVE(2011)

引用 4|浏览1
暂无评分
摘要
Introduction: Automated hand-held nerve conduction study (NCS) devices are being marketed for use in the diagnosis of lumbosacral radiculopathy (LSR). In this study we compared the specificity and sensitivity of a hand-held NCS device for the detection of LSR with standard electrodiagnostic study (EDX). Methods: Fifty patients referred to a tertiary referral electromyography (EMG) laboratory for testing of predominantly unilateral leg symptoms (weakness, sensory complaints, and/or pain) were included in the investigation. Twenty-five normal 'control' subjects were later recruited to calculate the specificity of the automated protocol. All patients underwent standard EDX and automated testing. Results: Raw NCS data were comparable for both techniques; however, computer-generated interpretations delivered by the automated device showed high sensitivity with low specificity (i.e., many false positives) in both symptomatic patients and normal controls. Conclusions: The automated device accurately recorded raw data, but the interpretations provided were overly sensitive and lacked the specificity necessary for a screening or diagnostic examination. Muscle Nerve 43: 9-13, 2011
更多
查看译文
关键词
automated nerve conduction studies, electromyography (EMG), radiculopathy, lumbosacral