Nstx Far Infrared Tangential Interferometer/Polarimeter Electronics Upgrade

REVIEW OF SCIENTIFIC INSTRUMENTS(2010)

引用 5|浏览1
暂无评分
摘要
New electronics for the multichannel far infrared tangential interferometer/polarimeter system employed on the National Spherical Torus Experiment (NSTX) have greatly extended its capability to monitor high frequency density fluctuations. Such measurements are essential in understanding transport physics issues in NSTX as well as for the coming ITER device. The electronics, which were previously limited to similar to 250 kHz, have been upgraded with a video bandwidth that extends to 4 MHz when operating as an interferometry-only configuration, and to similar to 500 kHz when operating as a simultaneous interferometer/polarimeter system. Experimental details and test results of the new electronics are presented. (C) 2010 American Institute of Physics. [doi:10.1063/1.3485103]