Image quality and radiation exposure in 320-row temporal bone computed tomography

DENTOMAXILLOFACIAL RADIOLOGY(2010)

引用 19|浏览5
暂无评分
摘要
Objectives: The aim was to define image quality and radiation exposure in the recently introduced 320-row CT of the temporal bone (tb) in comparison to a 16-row tb CT. Methods: A cadaveric head phantom was used for repeated tb volume CT studies (80120 kV, 25-150 mAs), performed in a 320-row scanner (single rotation, 0.5 mm slice thickness, kernel FC 51) in comparison to 16-row helical CT using standard acquisition parameters (SAP) of 120 kV and 75 mAs (kernel FC 53). Qualitative image evaluation was performed by two radiologists using a 5-point visual analogue scale. Image noise (D(SD)) was determined by region of interest (ROI) based measurements in cadaveric as well as water phantom studies. Dosimetric measurements of the effective dose (ED) and organ dose (OD) of the lens were performed. Results: Image quality of 320-row tb CT was equivalent to 16-row CT for SAP scans, resulting in image noise levels (D(SD) 16-/320-row) of 109/237 and 206/446 for air and bone respectively. D(SD) differences were predominantly (>90%) attributable to the different kernels available for tb studies in 16- and 320-row CT. Radiation exposure for 16-/320-row SAP scans amounted to 0.36/0.30 mSv (ED) and 10.0/8.4 mGy (lens dose). Conclusion: 320-row volume acquisition in tb CT delivers equivalent image quality to 16-row CT while decreasing radiation exposure figures by one sixth. Image noise increase in 320-row CT is negligible with respect to image quality. Dentomaxillofacial Radiology (2010) 39, 199-206. doi: 10.1259/dmfr/14139527
更多
查看译文
关键词
cone beam computed tomography,tomography, X-ray computed,temporal bone,petrous bone,radiation dosage
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要