Open MRI Magnet With Iron Rings Correcting the Lorentz Force and Field Quality

Applied Superconductivity, IEEE Transactions(2014)

引用 7|浏览25
暂无评分
摘要
A magnetic resonance imaging (MRI) superconducting magnet with openness can reduce claustrophobia of a patient. To fabricate an open MRI magnet, a new magnet configuration with iron rings and pole plates balancing the force between the coil and the yoke, and correcting the field homogeneity has been proposed. An MRI magnet with the center field of 0.7 T generated by split-pair coils and yokes is fabricated. The coils are contained in the cryostat with separated vessels connected with two pipes. The coils are separated with a room temperature gap of 480 mm and the diameter spherical volume (DSV) of 300 mm in which the homogeneity is 3 ppm in peak-to-peak. The cryogenic system is cooled by one GM cryo-cooler to realize the zero boiling off liquid helium. In the paper, the detailed design, fabrication, and test will be reported.
更多
查看译文
关键词
biomedical mri,biomedical equipment,cryogenics,superconducting magnets,gm cryo-cooler,lorentz force,claustrophobia,cryogenic system,diameter spherical volume,field quality,iron rings,magnet configuration,magnetic flux density 0.7 t,magnetic resonance imaging superconducting magnet,open mri magnet,pole plates,split-pair coils,temperature 293 k to 298 k,zero boiling off liquid helium,iron-rings balancing lorentz force,open magnetic resonance imaging (mri),superconducting magnet,magnetic resonance imaging,magnetic fields,helium,iron