Imaging of Magnetic Nanoparticles using second harmonic signals

Magnetics, IEEE Transactions(2015)

引用 4|浏览11
暂无评分
摘要
Magnetic nanoparticle imaging (MPI) utilizes the non-linear magnetization response M for the detection of superparamagnetic iron oxide nanoparticles (MNPs). When an external ac magnetic field is applied to the NMPs, some harmonic responses such as second, third and higher one arise. We propose and demonstrate that the use of the second harmonic response is the most effective technique to enhance the signal. The advantage of using the second harmonic response is that the response can be measured even in a small ac field. A two-dimensional (2-D) MPI system using this technique was realized and its performance was evaluated. The MPI system has no mechanical parts but instead magnetically scans an MNP phantom by changing the dc bias field in two directions. The cosine component of the signal is detected by a lock-in-amplifier and differentiated in one direction. The image of the phantom is then reconstructed using the d(cosθ)/dz data. As a result, the position of the phantom may be identified down to the iron content of 9 μg.
更多
查看译文
关键词
squid,magnetic nano-particles,second harmonic,iron,harmonic analysis,magnetic resonance imaging