Sensitive, fast-responding passive electrostatic radon monitor

R H Griffin,A Kochermin,N G Tarr,Henry D Mcintosh,Hui Ding, J P Weber, Renato A F Falcomer

Limerick(2011)

引用 2|浏览5
暂无评分
摘要
A second generation fast-responding passive radon detector using electrostatic concentration and enhanced readout electronics has been designed, built and tested. This detector utilizes the custom α-detecting IC called αRAM. In order to assess the response time, simulations were performed to analyze the turn-on transient of the detector. In an ambient with constant radon concentration, simulations indicate 1.44 hours will be required for the detector to reach 90% of its maximum count rate. This result is consistent with experimental findings using both first and second generation passive radon detectors.
更多
查看译文
关键词
electrostatic devices,gas sensors,integrated circuits,radon,random-access storage,αram,ic,fast-responding passive electrostatic radon monitor,readout electronics,prototypes,detectors,mathematical model