Non-metallic VEP stimulator without interference in CT scanning for neurosurgical operation

Shinji Fukuma,Yoshinori Kawakami, Ryuhei Kitai, Kousuke Awara,Tetsuma Sakurai, Atsushi Okuda

London(2006)

引用 0|浏览5
暂无评分
摘要
Non-metallic flash-type stimulators for monitoring visual evoked potential, named as the Fiber-Eyes, are proposed. Conventional VEP stimulators are comprised of LEDs and the print circuits, i.e. metallic materials, inside the stimulator. Proposed Fiber-Eyes are made from plastic optical fibers, silicone rubber, and external light source. Thus the Fiber-Eyes can be used together with CT scanning without interference in the image due to not use of the metallic parts during neurosurgical operations. They do not also radiate heat because the flash light for stimulus is supplied through the fibers, and thus it can avoid desiccation of the bulbar conjunctiva of patient in the operation.
更多
查看译文
关键词
biomedical engineering,computerised tomography,neurophysiology,surgery,visual evoked potentials,ct scan,fiber-eyes,flash type stimulators,interference,light emitting diodes,neurosurgical operation,nonmetallic vep stimulator,plastic optical fibers,silicone rubber,visual evoked potential,glass,computed tomography,current transformers,plastic optical fiber,optical fibers