Robotic system for MRI-guided stereotactic neurosurgery.

IEEE Trans. Biomed. Engineering(2015)

引用 112|浏览30
暂无评分
摘要
Stereotaxy is a neurosurgical technique that can take several hours to reach a specific target, typically utilizing a mechanical frame and guided by preoperative imaging. An error in any one of the numerous steps or deviations of the target anatomy from the preoperative plan such as brain shift (up to mm), may affect the targeting accuracy and thus the treatment effectiveness. Moreover, because the procedure is typically performed through a small burr hole opening in the skull that prevents tissue visualization, the intervention is basically “blind” for the operator with limited means of intraoperative confirmation that may result in reduced accuracy and safety. The presented system is intended to address the clinical needs for enhanced efficiency, accuracy, and safety of image-guided stereotactic neurosurgery for deep brain stimulation lead placement. The study describes a magnetic resonance imaging (MRI)-guided, robotically actuated stereotactic neural intervention system for deep brain stimulation procedure, which offers the potential of reducing procedure duration while improving targeting accuracy and enhancing safety. This is achieved through simultaneous robotic manipulation of the instrument and interactively updated in situ MRI guidance that enables visualization of the anatomy and interventional instrument. During simultaneous actuation and imaging, the system has demonstrated less than 15% signal-to-noise ratio variation and less than 0.20 geometric distortion artifact without affecting the imaging usability to visualize and guide the procedure. Optical tracking and MRI phantom experiments streamline the clinical workflow of the prototype system, corroborating targeting accuracy with three-ax- s root mean square error 1.38 ± 0.45 mm in tip position and 2.03 ± 0.58° in insertion angle.
更多
查看译文
关键词
medical robotics,image-guided therapy,magnetic resonance imaging (mri)-compatible robotics,mri-guided stereotactic neurosurgery robotic system,enhanced mri-guided stereotactic neurosurgery efficiency,small burr hole opening,neurosurgical technique,deep brain stimulation,neurophysiology,anatomy visualization,stereotaxy,mri-guided stereotactic neurosurgery safety,reducing procedure duration,targeting accuracy improvement,prototypes,imaging usability,in situ mri guidance,stereotactic neurosurgery,robot-assisted surgery,optical tracking,simultaneous robotic manipulation,biomedical mri,magnetic resonance imaging,stereotactic neural intervention system,signal-to-noise ratio variation,preoperative imaging,brain,mri phantom,interventional instrument,brain shift,tissue visualization,phantoms,surgery,mri-compatible robotics,mri-guided stereotactic neurosurgery accuracy,treatment effectiveness,accuracy,neurosurgery,robots
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要