A new method for high-energy pulsed Gamma measurement within intense background X-rays

CHINESE PHYSICS B(2010)

引用 3|浏览8
暂无评分
摘要
The accelerator-generating 6.13 MeV pulsed Gamma by (19)F(p, alpha gamma)(16)O reaction usually synchronizes with an intense bremsstrahlung x-ray which has a maximum energy of 1 MeV. This paper proposes a new method, named the scattering and absorbing method, to diagnose the 6.13 MeV Gamma. This method includes two parts: the detector and a scatterer placed in front of the detector. The detector converts the Gamma to electrons and then collects the electrons by a scintillator. In order to restrain the interference of the low-energy background, the scintillator collects the electrons at a small angle. The scintillator is wrapped with electro-absorbing material to absorb the low-energy electrons generated by background x-rays. The theoretical sensitivity ratio of 6.13 MeV Gamma to 1 MeV x-rays is greater than 150. The scatterer is a pretreatment tool to scatter some background x-rays away from the radial beam before they enter the detector. By varying the length, the scatterer can reduce the background x-rays to an acceptable level for the detector.
更多
查看译文
关键词
high energy pulsed Gamma,background,scattering,absorbing