Measurement of the gamma-ray sensitivity and signal-to-noise ratio of a new scattered-electron detector.

REVIEW OF SCIENTIFIC INSTRUMENTS(2008)

引用 2|浏览2
暂无评分
摘要
A new detector suitable for measuring high intensity pulsed gamma-ray sources and based upon scattered-electron method is proposed. The detector has a relatively flat energy response in the range of 0.4-5 MeV and works in current mode. The performances of the detector under several conditions were studied by Monte Carlo simulation using the MCNP code. A comparison between calculations and measurements performed using the 1.25 MeV line of Co-60 is also addressed. The experimental signal produced by the detector was thus studied and decomposed into its main components in order to establish the signal-to-noise ratio (SNR). The latter is compared to SNR calculated for other type of detectors. c 2008 American Institute of Physics. [DOI: 10.1063/1.2976670]
更多
查看译文
关键词
gamma ray,monte carlo method,monte carlo simulation,remote sensing,signal to noise ratio