Pilot study in the identification of stability values for determining immediate and early loading of implants.

The Journal of oral implantology(2009)

引用 15|浏览2
暂无评分
摘要
This pilot study assessed the impact of implant stability criteria on implant loading time. Insertion torque (IT) and baseline implant stability quotients (ISQ) from resonance frequency analysis were recorded for 41 tapered-screw implants placed in the mandibles of 20 patients. Immediate (IL), early (EL) (6 weeks postoperative), or delayed (DL) (12 weeks postoperative) prosthetic loading was performed based on results and study criteria. Postoperative ISQ values were also recorded at 6 time intervals for the IL and EL groups and after healing for the DL group. All implants were definitively restored after 6 weeks of provisionalization. There were no failures, complications, or stability differences based on loading time. At week 12 of follow-up, ISQ values were slightly higher for IL implants compared with EL implants. IL and EL implants showed significantly higher cumulative ISQ values compared with DL implants. High IT and baseline ISQ values in all groups corresponded to high implant survival. Postoperative ISQ values in the IL and EL groups were less certain indicators of implant survival because of fluctuations relative to baseline values. Small sample size and inequitable patient distribution across groups skewed results. More research is needed before definitive results can be drawn. IL and EL were safely performed within the IT and ISQ ranges in this study, but it is unknown whether EL criteria would have also sufficed for IL.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要