Operation Of A High-Tc Dc-Squid-Gradiometer On A Non-Metallic Pulse Tube Refrigerator

8TH EUROPEAN CONFERENCE ON APPLIED SUPERCONDUCTIVITY (EUCAS'07)(2008)

引用 7|浏览4
暂无评分
摘要
A planar high-T-c DC-SQUID gradiometer has been operated with a specially developed low noise pulse tube refrigerator. The cold finger of the refrigerator consists only of non-metallic and non-magnetic materials. The high sensitive SQUID-gradiometer with a base length of 4 mm has a field gradient resolution of about 3 pT/(cm Hz(1/2)) at 1 Hz and 0.7 pT/(cm Hz(1/2)) in the white noise region in magnetically shielded environment. During the operation of both, the sensor and the refrigerator, the noise generated by the cryocooler is below the noise level of the SQUID-gradiometer. We demonstrate the potential of this nonmetallic pulse tube refrigerator by measuring the magnetic field originating from a human heart (magnetocardiogram), without additional suppression of the intrinsic refrigerator noise.
更多
查看译文
关键词
magnetic materials,white noise,magnetic field