LC-MS analysis of phospholipids and lysophospholipids in human bronchoalveolar lavage fluid.

Journal of Chromatography B(2005)

引用 63|浏览2
暂无评分
摘要
A reversed phase HPLC method was developed for the simultaneous analysis of different phospholipids and lysophospholipids in human bronchoalveolar lavage fluid (BALF). Separation was achieved using a pellicular C8 column at elevated temperatures with an increasing gradient of acetonitrile containing 0.1% formic acid. Detection was carried out by electrospray ionization ion-trap mass spectrometry. Calibration graphs for selected phospholipids (phosphatidylcholine, phosphatidylethanolamine, phosphatidylglycerol and lysophosphatidylcholine) showed linearity up to 50ng allowing quantitative determinations. Identification of the individual species within each class was possible with tandem mass spectrometry. Analysis of BALF phospholipids was performed after liquid/liquid extraction with a mixture of chloroform/methanol/acetic acid. Recoveries ranged from 69 to 97% with standard deviations of less than 6%. The limit of detection varied slightly between different classes but was in the range 0.05–0.25ng total injected amount.
更多
查看译文
关键词
CID,TIC,EIC,LC–MS,COPD,BAL,BALF,PL,LPL,PC,PE,PG,LPC