Monte Carlo simulation of glandular dose in a dedicated breast CT system

CHINESE PHYSICS C(2012)

引用 3|浏览18
暂无评分
摘要
A dedicated breast CT system (DBCT) is a new method for breast cancer detection proposed in recent years. In this paper, the glandular dose in the DBCT is simulated using the Monte Carlo method. The phantom shape is half ellipsoid, and a series of phantoms with different sizes, shapes and compositions were constructed. In order to optimize the spectra, monoenergy X-ray beams of 5-80 keV were used in simulation. The dose distribution of abreast phantom was studied: a higher energy beam generated more uniform distribution, and the outer parts got more dose than the inner parts. For polyenergtic spectra, four spectra of Al filters with different thicknesses were simulated, and the polyenergtic glandular dose was calculated as a spectral weighted combination of the monoenergetic dose.
更多
查看译文
关键词
dedicated breast CT,Monte Carlo simulation,glandular dose