Cavity ring-down technique for measurement of reflectivity of high reflectivity mirrors with high accuracy

Pramana(2011)

引用 9|浏览3
暂无评分
摘要
A simple, accurate and reliable method for measuring the reflectivity of lasergrade mirrors ( R > 99.5%) based on cavity ring-down (CRD) technique has been successfully demonstrated in our laboratory using a pulsed Nd:YAG laser. A fast photomultiplier tube with an oscilloscope was used to detect and analyse the CRD signal. The cavity decay times were measured for three cavities formed by a combination of three mirror pairs. The absolute reflectivities R 1 , R 2 , R 3 were determined to be 99.94%, 99.63%, 99.52% at normal incidence. The reflectivity of mirrors is measured to an accuracy of 0.01%.
更多
查看译文
关键词
Cavity ring-down method,reflectivity measurement,optical resonator