A patch in the pectoral position lowers defibrillation threshold.

PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY(1997)

引用 7|浏览14
暂无评分
摘要
Implantable pacemaker cardioverter defibrillators are now available with biphasic waveforms, which have been shown to markedly improve defibrillation thresholds (DFTs). However, in a number of patients the DFT remains high. Also, DFT may increase after implantation, especially if antiarrhythmic drugs are added. We report on the use of a subcutaneous patch in the pectoral position in 15 patients receiving a transvenous defibrillator as a method of easily reducing the DFT. A 660-mm(2) patch electrode was placed beneath the generator in a pocket created on the pectoral fascia. The energy required for defibrillation was lowered by 56% on average, and the system impedance was lowered by a mean of 25%. This maneuver allowed all patients to undergo a successful implant with adequate safety margin.