Central corneal thickness measurements using Orbscan II scanning slit topography, noncontact specular microscopy, and ultrasonic pachymetry in eyes with keratoconus.

Cornea(2005)

引用 52|浏览6
暂无评分
摘要
PURPOSE:To compare corneal thickness measurements using Orbscan II scanning slit topography, Topcon SP-2000P noncontact specular microscopy, and ultrasonic pachymetry in eyes with keratoconus. METHODS:Central corneal thickness was measured in 22 eyes with keratoconus. Eyes with apparent corneal opacity were excluded. Scanning slit topography, noncontact specular microscopy, and ultrasonic pachymetry were used in this sequence. The acoustic equivalent correlation factor (0.92) was used for Orbscan readings. RESULTS:Three devices gave significantly different corneal thickness readings (P < 0.001, repeated-measure analysis of variance). Measurements with Orbscan scanning slit topography (449.5 +/- 43.2 [SD] mum) were significantly smaller than those of ultrasonic pachymetry (485.0 +/- 29.3 microm; P < 0.001, Tukey multiple comparison) and SP-2000P noncontact specular microscopy (476.7 +/- 28.3 microm; P = 0.002). There were significant linear correlations between ultrasonic pachymetry and scanning slit topography (Pearson correlation coefficient r = 0.741, P < 0.001), between scanning slit topography and noncontact specular microscopy (r = 0.880, P < 0.001), and between noncontact specular microscopy and ultrasonic pachymetry (r = 0.811, P < 0.001). CONCLUSION:In eyes with keratoconus, Orbscan II scanning slit topography system gave significantly smaller corneal thickness readings than the other 2 devices. Measurements taken by noncontact specular microscopy and ultrasonic pachymetry were comparable. Three devices showed significant linear correlations with one another.
更多
查看译文
关键词
pachymetry,corneal thickness,keratoconus,Orbscan scanning slit topography,noncontact specular microscopy
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要