The novel sensitive and high throughput determination of cefepime in mouse plasma by SCX-LC/MS/MS method following off-line μElution 96-well solid-phase extraction to support systemic antibiotic programs

Journal of Chromatography B(2010)

引用 20|浏览5
暂无评分
摘要
A sensitive and high throughput off-line μElution 96-well solid-phase extraction (SPE) followed by strong cation exchange (SCX) liquid chromatography with tandem mass spectrometry (LC/MS/MS) quantification for determination of cefepime has been developed and validated in mouse plasma. Using the chemical analog, ceftazidime as an internal standard (IS), the linear range of the method for the determination of cefepime in mouse plasma was 4–2048ng/mL with the lower limit of quantitation level (LLOQ) of 4ng/mL. The inter- and intra-assay precision and accuracy of the method were below 9.05% and ranged from 95.6 to 113%, respectively, determined by quality control (QC) samples at five concentration levels including LLOQ. After μElution SPE, 71.1% of cefepime was recovered. The application of the validated assay for the determination of cefepime in mouse pharmacokinetics (PK) samples after intravenous (IV) and subcutaneous (SC) doses was demonstrated.
更多
查看译文
关键词
Cefepime,μElution SPE,SCX,LC/MS/MS,Mouse plasma,Pharmacokinetics