Optimization of preparation procedure of LiF:Mg,Cu,Si TLD for improving the reusability

Radiation Measurements(2010)

引用 8|浏览18
暂无评分
摘要
Several thermal treatments in the temperature range from 270 °C to 320 °C (each of 10 min) were tested as a final preparation procedure of LiF:Mg,Cu,Si to improve the protocol of TL readout with less residual signal for the LiF:Mg,Cu,Si TLD. This high sensitivity LiF:Mg,Cu,Si TLD exhibited thermal stability much better than that of the well known LiF:Mg,Cu,P. For LiF:Mg,Cu,Si, a readout temperature up to 300 °C did not affect the TL sensitivity and glow curve structure for 12 cycles of exposure and readout following an initial thermal treatment at 295 °C for 10 min. The residual TL signal also remained negligible.
更多
查看译文
关键词
LiF:Mg,Cu,Si,Thermoluminescence,TLD,Residual signal