Determination of metformin in plasma by high-performance liquid chromatography.
Journal of Chromatography B: Biomedical Sciences and Applications(1992)
摘要
A high-performance liquid chromatographic method for the determination of metformin, an oral antidiabetic agent, in plasma is described. Plasma samples containing the internal standard, phenformin, are eluted through Amprep extraction columns before injection into the chromatographic column, packed with μBondapak phenyl. The eluent is monitored at 236 nm. At a mobile phase flow-rate of 1.35 ml/min, the retention times of metformin and phenformin are 2.8 and 5.6 min, respectively. The intra-day coefficients of variation are 1.5 and 4.3% at metformin concentrations of 0.05 and 1 mg/l, respectively.
更多查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要