Development of a new two-dimensional position-sensitive detection based on resistive charge division and using CdTe detectors for a high-resolution semiconductor-based PET scanning

Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment(2011)

引用 5|浏览15
暂无评分
摘要
A new position-sensitive detection technique for semiconductor detector fabrication was developed for a high-spatial-resolution PET scanner with CdTe detectors. We constructed a two-dimensional sensitive CdTe detector with a size of 20×9.5×1.0mm3, of which one side had 1mm wide resistive strip anodes. Using our method, we succeeded in obtaining a linear presentation of a two-dimensional detection image from a distorted detection image. We showed that the detection positioning within a width of 1.2mm for the x-axis was obtained with a confidence level of 0.98, and a spatial resolution of approximately 1mm FWHM was obtained for the y-position. Our newly developed position-sensitive detector will enable the construction of a high-performance PET scanner due to the detector's high spatial resolution and good sensitivity; such a PET scanner is now under construction in our laboratory.
更多
查看译文
关键词
Position-sensitive detector,Semiconductor PET