Development Of An External Faraday Cup For Beam Current Measurements

Kye-Ryung Kim,Myung-Hwan Jung, Se Jin Ra, Seok Ki Lee, Sun Taek Hwang

JOURNAL OF THE KOREAN PHYSICAL SOCIETY(2010)

引用 7|浏览2
暂无评分
摘要
In general, beam current measurements are very important for many kinds of experiments using highly energetic particle beams at accelerators, such as cyclotrons, linacs, etc. The Faraday cup is known to be one of the most popular beam current measurement tools. We developed an external Faraday cup to measure the beam current at a dedicated beam line for low-flux experiments installed at the MC-50 cyclotron of Korea Institute of Radiological & Medical Sciences (KIRAMS). It was designed for external beam current measurements and is composed of a vacuum chamber, an entrance window, a collimator, a electrostatic suppressor ring, and a cup. The window is made of 75-um-thick Kapton film, and the diameter of the collimator is 10 mm or 20 mm. The ring and the cup has 5-cm inner diameters, and the thickness of the bottom of the cup is 2 cm, which is enough to absorb the total proton energy up to 45 MeV. Using this external Faraday cup, we measured the beam current from the cyclotron, and we compared measured flux to the results from film dosimetry using GAF films.