Low-Noise Scintillation Detectors With A P-47 Thin-Layer Screen For Electrons Of Several Kev
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS(1994)
摘要
The applicability of a low-noise scintillation detector (ScD) for the registration of electrons of several keV energy has been studied employing photomultipliers (PM) of different types and sizes. With the application of a sedimented P-47 scintillation screen, the values of the low-energy sensitivity limit and those of the light conversion coefficient were determined as about 2.7-4.7 keV and 2.8-6.6 photoelectrons/keV, respectively, for the set of PM's (Philips-Valvo XP 2020, Philips-Valvo XP 2052, Philips-Valvo XP 2972, EMI 9124a) studied.It is concluded that such scintillation detectors might be used advantageously as electron counters in the range of E > 5 keV. Applications below this kinetic energy value are also feasible when applying a floating acceleration of several kV to the ScD - a voltage much lower than the values required for Everhart-Thornley detectors.
更多查看译文
关键词
kinetic energy,scintillation detector