Lesiones producidas por la ablación con radiofrecuencia del istmo cavotricuspídeo en un modelo experimental

Revista Española de Cardiología(2003)

引用 1|浏览17
暂无评分
摘要
Introduction and objectives. Experimental studies have shown that deeper and wider lesions (up to 10 mm long or deep) can be safely created using an 8 mm or irrigated tip catheter for ablation to treat atrial flutter. However, potential damage to the tricuspid valve or inferior cava vein has not been systematically evaluated. Patients and method. The cavotricuspid isthmus was ablated in 26 pigs (body weight 26-52 kg), with a total of 187 radiofrequency pulses. Standard 4 mm, 8 mm and irrigated tip catheters were used at random. For each ablation, energy, impedance and temperature were recorded continuously. Results. The lesions were larger with irrigated tip and 8-mm catheters than with standard ones. In 7 animals (1 with an irrigated tip, 4 with an 8-mm, and 2 with a standard tip) the tricuspid valve was damaged. The tricuspid valve was severely damaged in 3 pigs and lesions were moderate in 4. In animals with tricuspid valve lesions, maximal energy was higher (59 +/- 27 vs. 51 +/- 24 W; p = 0,03) and higher temperatures were reached (63 4 vs. 55 +/- 11 degreesC; p < 0.001). Low energy pulses measured before ablation were also more intense in animals in which damage was produced (0.55 +/- 0.24 vs. 0.35 +/- 0.29; p = 0.001), indicating greater contact pressure. Conclusions. The tricuspid valve may be severely damaged during the ablation of the cavotricuspid isthmus for atrial flutter: damage was seen most often with high energy pulses and with 8-mm catheters, but can also occur with usual energy levels and standard catheters. To minimize damage this technique should not be used from the inside of the right ventricle just above the tricuspid valve.
更多
查看译文
关键词
Radiofrecuencia,Ablación,Aleteo auricular
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要