Optimal configuration of capillary electrophoresis microchip with expansion chamber in separation channel.

Journal of Chromatography A(2006)

引用 15|浏览5
暂无评分
摘要
This study develops a novel capillary electrophoresis (CE) microfluidic device featuring a conventional cross-form injection system and an expansion chamber located at the inlet of the separation channel. The combined injection system/expansion chamber arrangement is designed to deliver a high-quality sample band into the separation channel such that the detection performance of the device is enhanced. Numerical simulations are performed to investigate the electrokinetic transport processes in the microfluidic device and to establish the optimal configuration of the expansion chamber. The results indicate that an expansion chamber with an expansion ratio of 2.5 and an expansion length of 500μm delivers a sample plug with the correct shape and orientation. With this particular configuration, the peak intensities of the sample are sharp and clearly distinguishable in the detection region of the separation channel. Therefore, this configuration is well suited for capillary electrophoresis applications which require a highly sensitive resolution of the sample plug. The novel CE microfluidic device developed in this study has an exciting potential for use in high-performance, high-throughput chemical analysis applications and in many other applications throughout the field of micro-total-analysis-systems.
更多
查看译文
关键词
Injection system,Expansion chamber,Capillary electrophoresis,Micro-total-analysis-systems