Elevated defibrillation thresholds in patients undergoing biventricular defibrillator implantation: Incidence and predictors

Heart Rhythm(2006)

引用 53|浏览10
暂无评分
摘要
Conclusion Patients with a biventricular ICD had a 12% incidence of elevated DFT in our sequential patient cohort. QRS duration prior to biventricular ICD placement is the most powerful predictor of patients at risk for an elevated DFT. An elevated DFT does not have an impact on mortality, perhaps because of successful implementation of system modifications to ensure an adequate defibrillation safety margin.
更多
查看译文
关键词
Biventricular implantable cardioverter-defibrillator,Defibrillation threshold,Defibrillation efficacy