Development and validation of a LC/MS/MS method for quantification of nobiliside A in rat plasma.

Journal of Chromatography B(2009)

引用 5|浏览7
暂无评分
摘要
A LC/MS/MS method was developed and validated for determination of nobiliside A in rat plasma. Analyses were separated by an Agella Venusil ASB-C18 column and isocratic elution with methanol:water (80:20, v/v) as a mobile phase. A tandem mass spectrometer was used as a detector for quantitative analysis. Calibration curves (R2>0.999) in spiked plasma were linear over the concentration range of 50–5000ng/mL. The overall accuracy of this method was 93–104% for nobiliside A. Within-day and between-day variability was both less than 9% in plasma. The data indicate that the LC/MS/MS method is an effective method for the pharmacokinetics study of nobiliside A in rat plasma.
更多
查看译文
关键词
Nobiliside A,Plasma analysis,LC/MS/MS