Evaluation Of Respiratory Inductive Plethysmography In Controlled Ventilation - Measurement Of Tidal Volume And Peep-Induced Changes Of End-Expiratory Lung Volume

CHEST(1998)

引用 45|浏览2
暂无评分
摘要
Study objective: To determine the accuracy of respiratory inductive plethysmography (RIP) with a respiratory monitor (Respitrace Plus; NIMS Inc., Miami) operating in the DC-mode for the measurement of tidal volumes (VT) and positive end-expiratory pressure (PEEP)-induced changes of end-expiratory lung volume (Delta EELV) in patients with normal pulmonary function, acute lung injury (ALI), and COPD during volume-controlled ventilation.Design: Prospective comparison of RIP with pneumotachography (PT) for assessment of VT and with multibreath nitrogen washout procedure (N-2,WO) for determination of Delta EELV as reference methods.Setting: Mixed ICU at a university hospital.Patients: Thirty-one sedated and paralyzed patients: 12 patients with normal pulmonary function mechanically ventilated after major surgery, 10 patients with respiratory failure due to ALI, and 9 patients with a known history of COPD ventilated after surgery or because of respiratory failure due to bronchopulmonary infection.Interventions;: Stepwise increase of PEEP from 0 to 5 to 10 cm H-2,O and reduction to 0 cm H-2,O again. On each PEEP level, N-2,WO was performed.Measurements and main results: The baseline drift of RIP averaged 25.4+/-29.1 mL/min but changed over a wide range even in single patient measurements. Determination of VT for Single minutes revealed that 66.5% and 90.0% of all values were accurate within a range of +/-10% and +/-20%, respectively. The deviation for VT measurements between RIP and PT in patients with COPD was significantly (p<0.05) higher compared to patients with ALI or normal pulmonary function. The difference of Delta EELV between RIP and N2WO was 11.6+/-174.1 mt with correlation coefficients of 0.77 (postoperative and COPD patients) and 0.86 (ALI patients). However, just 25.8% and 46.2% were precise within +/-10% and +/-20%, respectively. Delta EELV determination in COPD patients differed more between RIP and N-2,WO than in the other groups, but this was not significant.Conclusion: In a mixed group of patients undergoing controlled ventilation, RIP using the Respitrace Plus monitor was not consistently precise enough for quantitative evaluation of VT and EELV when compared to our reference methods. This was most evident in patients with COPD. For long-term volume measurements, a better control of the baseline drift of RIP should be achieved.
更多
查看译文
关键词
acute lung injury, COPD, end-expiratory lung volume, respiratory inductive plethysmography, tidal, tidal volume
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要