THE HUMPHREY ADE SYSTEM: EVALUATION IN PAEDIATRIC USE
BRITISH JOURNAL OF ANAESTHESIA(1991)
摘要
The Humphrey ADE single lever parallel form breathing system was evaluated in 17 children whose lungs were ventilated mechanically, mean age 28 (range 9-58) months, weight 13.3 (8.3-18.0) kg and in 10 spontaneously breathing children, age 50.7 (18-99) months and weight 17.5 (10.9-24.3) kg. During controlled ventilation in the E mode, the ADE behaved in a manner similar to the Jackson-Rees modification of Ayre's T-piece with the fresh gas flows (FGF) set either according to weight or at a standard 3 litre min-1. After 5 min hand ventilation in the A mode with FGF 3 liter min-1 the mean end-tidal partial pressure of carbon dioxide was 5.3 (SD 0.8) kPa (range 4.1-7.7 kPa). Whilst breathing spontaneously in the A mode, no patient experienced rebreathing at FGF 3 litre min-1; rebreathing started at 124 (31.6) ml kg-1 min-1. The ADE circuit performed satisfactorily in all three modes. During controlled ventilation in the E mode, FGF should be adjusted according to the standard formulae recommended for the T-piece circuit. During hand ventilation in the A mode, an FGF greater than 3 litre min-1 should be used.
更多查看译文
关键词
ANESTHESIA,PEDIATRIC,EQUIPMENT,BREATHING SYSTEMS,HUMPHREY ADE SYSTEM