NaI (Tl) Calorimeter Calibration and Simulation for Coulomb Sum Rule Experiment in Hall-A at Jefferson Lab

CHINESE PHYSICS C(2011)

引用 0|浏览15
暂无评分
摘要
A precision measurment of inclusive electron scattering cross sections is carried out at Jefferson Lab in the quasi-elastic region for He-4, C-12, Fe-56 and Pb-208 targets. The longitudinal (R-L) and transverse (R-T) response functions of the nucleon need to be extracted precisely in the momentum transfer range 0.55 GeV/c <= vertical bar q vertical bar <= 1.0 GeV/c. To achieve the above goal, a NaI (Tl) calorimeter is used to distinguish good electrons from background, including pions and low energy electrons rescattered from the walls of the spectrometer magnets. Due to a large set of kinematics and changes in HV settings, a number of calibrations are performed for the NaI (Tl) detector. Corrections for a few blocks of NaI (Tl) with bad or no signal are applied. The resolution of the NaI (Tl) detector after calibration reached delta E/root E approximate to 3% at E=1 GeV. The performance of the NaI (Tl) detector is compared with a simulation. The good calibration and background analysis for the NaI(Tl) detector are very important for the reduction of the systematic error of cross sections and the separation of R-L and R-T.
更多
查看译文
关键词
Coulomb sum rule,NaI (Tl),calibration,resolution,GEANT4 simulation